Portable Total Hydrocarbon Analyzer (0-50,000 ppm)

Classification:

The portable total hydrocarbon analyzer is an analytical instrument designed for the rapid detection of total hydrocarbon levels in the environment. It employs advanced Micro-FID technology and sophisticated data-processing algorithms, enabling accurate and swift measurement of total hydrocarbon concentrations in air. Featuring user-friendly operation, high sensitivity, and real-time data display, this instrument is widely applicable in fields such as environmental monitoring, industrial safety, and emergency response systems.

Key words:

Product Description


The portable total hydrocarbon analyzer is an analytical instrument designed for the rapid detection of total hydrocarbon levels in the environment. It employs advanced Micro-FID technology and sophisticated data-processing algorithms, enabling accurate and swift measurement of total hydrocarbon concentrations in air. Featuring user-friendly operation, high sensitivity, and real-time data display, this instrument is widely applicable in fields such as environmental monitoring, industrial safety, and emergency response systems.

Product advantages


Linear range: 0–50,000 ppm (0–100,000 ppm optional)

Operating temperature: - 20°C to +50°C

Resolution: 0.01ppm

Transmission method: Supports multiple control methods (Bluetooth, Wi-Fi, 5G)

Good portability: Small in size, lightweight, and easy to carry around.

High level of protection: IP66

Application areas


  • LDAR Project: VOCs Leak Detection
  • Natural Gas and Oil Pipeline Leak Detection (CH₄)
  • Breathing Valve and Safety Valve Safety Assessment
  • Soil Gas Safety Testing
  • Valve Low-Leakage Certification Testing
  • Environmental Protection Enforcement Inspection

Product Parameters


Project

Technical specifications

Linear range

0–50,000 ppm (0–100,000 ppm optional)

Detector

FID, PID (optional)

Limit of detection

0.1 ppm (CH₄)

Data Transmission

Bluetooth wireless transmission or Wi-Fi

Response time

<3.5s

Weight

<3kg

Linear error

±1.5%

Accuracy

±10%

Hydrogen usage time

The high-pressure gas cylinder can operate continuously for more than 8 hours.
Optional solid-state hydrogen storage tanks provide up to 20 hours of continuous use.